EL3020 ABB Continuous Gas Analyzers

Dynamic response
Warm-up time
Approx. 2 hours
90% response time
T90 ≤ 3 sec for measurement cell length = 220 mm, sample
gas flow = 60 l/h and electronic T90 time = 0 sec

Category: Tag:
Whatsapp:+86 13313705507
WeChat:+86 13313705507
E-mail:geabbdcs@gmail.com
Contacts:KEN HE

Description

EL3020 ABB Continuous Gas Analyzers

Stability
The following data apply only if all influence factors (e.g. flow rate, temperature, atmospheric pressure) are constant. They relate to the smallest measurement range.

Linearity deviation

≤ 1 % of span
Repeatability
≤ 0.5 % of span
Zero drift
≤ 1 % of span per week
Sensitivity drift
≤ 1 % of measured value per week
Output fluctuation (2 σ)
≤ 0.2 % of span at electronic T90 time (static/dynamic)
= 5/0 sec
Detection limit (4 σ)
≤ 0.4 % of span at electronic T90 time (static/dynamic)
= 5/0 sec

EL3020 ABB Continuous Gas Analyzers

EL3020 ABB Continuous Gas Analyzers

Measurement principle
Potentiometric measurement; zirconium dioxide cell for determination of the oxygen concentration in accordance with Nernst’s equation; reference gas: ambient air.The analyzer is used for the continuous measurement of oxygen in pure gases (N2, CO2, Ar). The measuring cell is catalytically inactivated to the extent that flammable carrier components in stoichiometric concentrations only negligibly reduce the oxygen value.

Sample component and measurement ranges
Sample component
Oxygen (O2)
Measurement range quantity and limits
2 measurement ranges
Measurement ranges are freely adjustable within the 0–1 ppm
to 0–250,000 ppm O2 range; they are factory-set to 0–1/
0–10 ppm O2.
The following measurement data refer to a measurement span of 100 ppm O2 with a regulated flow rate of 8 ± 0.2 l/h.

EL3020 ABB Continuous Gas Analyzers

EL3020 ABB Continuous Gas Analyzers

Notes
Both drawings show examples for the I/O modules equipment.
I/O module connection:
– The maximum capacity of terminals for stranded or solid conductors is 1 mm2
(17 AWG).
– The stranded conductor may be tinned on the tip or twisted  for simplified connection.
– When using wire end ferrules the total section should not exceed 1 mm2 , i.e. the maximum stranded conductor section is 0.5 mm2. The Weidmüller PZ 6/5 crimping tool must be used for crimping the ferrules. Functional scope of the Ethernet interface:
– Communication with configuration software ECT for gas analyzer configuration and software update.
– QAL3 data transfer if the QAL3 monitoring option is integrated in the gas analyzer.

EL3020 ABB Continuous Gas Analyzers PDF

EL3020

EL3020 ABB Continuous Gas Analyzers

…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦

Shipping:
1. Your item will be shipped to all countries via EMS, unless one country cannot be shipped via EMS. We will choose DHL, UPS, or FedEx. Please leave us your phone number.
2. We will ship the product to you on the same day or within 1-2 working days after receiving your payment. The shipping time depends on the country/region where you reside, the customs and habits of your country/region, and especially during peak hours.
3. Assuming (assuming) that the item requires payment of tariffs from your country, the buyer is obligated. We are not responsible for any taxes or customs fees in your country/region.

 

Return Policy:
1. Please contact us before returning the item.
2. The buyer should inform the seller of all returned items within 7 days after receiving the package. We reserve the right to reject any request to inform the seller for more than 7 days.
3. Returned products must have all original packaging and be in good condition.

 

Response policy:
If you are satisfied with the purchase, please leave us an active response.
2. If you have any questions about the outcome of the transaction or product, please contact us first and we will do our best to satisfy you!
3. Please do not reply with negative feedback before contacting us!

 

 

More models available:

DS200CTBAG1ADD

9905-263

5233-2089

IC8008SI228RD2

8237-1006

5233-2089

8446-1019

EASYGEN-3500-5/P2 8440-2145

9905-017

8200-314