ABB AO2000 Gas Analyzer Testing Methodology
Applicable Models: AO2000 Series (Software Version ≥3.0)
1. Pre-Test Preparation
1.1 Safety Compliance
- Verify grounding resistance ≤4Ω and insulation resistance ≥100MΩ (500V DC)
- Confirm purge gas supply (N₂/Ar) pressure at 2-5 bar with flow rate ≥20 L/min
- Wear ESD-protective gear when handling electronic modules
1.2 System Initialization
1. Power sequence: - Activate main breaker → Enable 24V DC control circuit → Start display unit 2. Warm-up duration: - IR detectors (Limas11/Uras14): ≥60 mins - Electrochemical cells (Magnos17): ≥30 mins 3. Perform auto-diagnostic via [System] → [Self-Test]
2. Calibration Testing
2.1 Manual Calibration (Single-Point)
| Step | Action | Parameter Range |
|---|---|---|
| 1 | Select target component (O₂/CO/NOx etc.) | Via [Calibration] → [Component Select] |
| 2 | Apply certified test gas | Concentration 80-100% of range |
| 3 | Stabilize flow at 1.5±0.2 L/min | Pressure drop ≤50 mbar |
| 4 | Initiate calibration | [Start Manual Cal] → Wait for “Stable” indicator |
| 5 | Validate results | Error ≤±1% FS (≤±2% for trace components) |
2.2 Automated Multi-Point Calibration
- Configure test sequence in [Auto-Cal Program]:
[Sequence_1] Gas1 = SpanGas (100% range) Gas2 = ZeroGas (N₂ 5.0) Gas3 = MidGas (50% range) Interval = 24h ; Default calibration cycle - Enable cross-sensitivity compensation for CO/CO₂ mixtures
3. Functional Verification
3.1 Dynamic Response Test
- Step change test (10→90% range):
- Response time (T90): ≤15s for NDIR, ≤30s for electrochemical
- Overshoot: ≤5% of final value
3.2 Limit Value Monitoring
- Set alarms at:
- Warning: 80% of range
- Critical: 95% of range
- Verify relay activation via [I/O Test] menu
4. Post-Test Procedures
4.1 Data Documentation
- Export calibration logs (CSV format) including:
Timestamp, Component, ExpectedValue, MeasuredValue, Deviation, OperatorID - Save configuration snapshot using [File] → [Export Config]
4.2 System Reset
- Purge sample lines for 10 mins with inert gas
- Deactivate test mode via [Operation] → [Normal Run]
AO2000 ABB pdf
♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦♦…♦
Shipping:
1. Your item will be shipped to all countries via EMS, unless one country cannot be shipped via EMS. We will choose DHL, UPS, or FedEx. Please leave us your phone number.
2. We will ship the product to you on the same day or within 1-2 working days after receiving your payment. The shipping time depends on the country/region where you reside, the customs and habits of your country/region, and especially during peak hours.
3. Assuming (assuming) that the item requires payment of tariffs from your country, the buyer is obligated. We are not responsible for any taxes or customs fees in your country/region.
Return Policy:
1. Please contact us before returning the item.
2. The buyer should inform the seller of all returned items within 7 days after receiving the package. We reserve the right to reject any request to inform the seller for more than 7 days.
3. Returned products must have all original packaging and be in good condition.
Response policy:
If you are satisfied with the purchase, please leave us an active response.
2. If you have any questions about the outcome of the transaction or product, please contact us first and we will do our best to satisfy you!
3. Please do not reply with negative feedback before contacting us!
More models available: